Cryotherapy Room Requirements: What to Check Before Your Chamber Arrives

By the Cryotherapy Service Repair Technical Team

Short answer

Before a cryotherapy chamber is delivered, confirm six things: the delivery path, the floor space, the power supply, how the heat leaves the room, room humidity below 40%, and network access. Nitrogen cryosaunas add two more: oxygen monitoring and ventilation, and safe nitrogen tank storage and delivery. Most installation problems we are called to fix come from one of these.

Download the room checklist (PDF) · Talk to us about your installation →

1. Delivery path

The chamber has to get from the truck to the room.

  • Measure every door, corridor, turn and elevator on the way in. Some chambers arrive as one piece and cannot be taken apart; the CTN eCabin, for example, ships as a single unit of about 65 × 35 × 94 in.
  • Check separate components too. CryoBuilt asks for doorways of at least 78 × 32 in along the route for its cooling unit.
  • Check the floor load. Chambers weigh from several hundred to well over 1,000 lb (the Artica Element is 1,200 lb).

2. Space in the room

  • Footprint plus clearance. Manufacturers specify an installation space larger than the chamber itself, for door swing, service access and airflow. Example: the Artica Element measures 76 × 36 in but needs 105 × 71 in of floor space; the NEO Luna Solo needs a room of at least 8 × 12 ft.
  • Ceiling height. Check against the installation height, and look for sprinklers, vents and lights directly above the unit.
  • Service access. Leave room for a technician to reach the back panels and the cooling unit. A chamber pushed tight into a corner costs more to repair.

3. Power

  • Single-phase or 3-phase. Newer self-contained chambers run on single-phase 208–240 V. Many first-generation chambers need 3-phase power, and some imported units need a transformer to match US voltage. 3-phase service is not available in every building.
  • A dedicated circuit sized to the manufacturer’s specification (for example 208–240 V, 30 A for the NEO Luna Solo).
  • A licensed electrician to install it, with any permit your city requires.

See the installation specs by model for each chamber’s power requirement.

4. Getting the heat out

Every electric chamber removes heat from the cabin and puts it somewhere.

  • Self-contained chambers release heat into the room. Plan room air conditioning for it; Artica, for example, specifies at least 12,000 BTU of room cooling for the Element. Some self-contained units put out so much heat that strong room AC is essential.
  • Split systems move the heat outside through a condenser on the roof or the ground. That means refrigerant line sets, wall or roof penetrations, roof access and usually a refrigeration contractor. Refrigerant work must be done by an EPA 608 certified technician.
  • Moving later. A split system has to be decommissioned and re-installed at the new location. A self-contained, plug-in unit can be disconnected and moved in hours.

5. Humidity and ventilation

Moisture is the main cause of avoidable repairs we see.

  • Keep the room below 40% relative humidity. Use a dehumidifier if needed. Artica specifies the same limit for the Element.
  • Dry the chamber completely every night. Moisture left inside freezes on fans, lifts and sensors.
  • Ventilate the room. Follow the manufacturer’s air-change requirement (Artica specifies at least 10 air changes per hour for the Element).
  • Keep the room temperature moderate. Artica specifies under 85 °F, ideally 60–70 °F.

6. Network

Many chambers support remote diagnostics and software updates. Provide a reliable Ethernet or Wi-Fi connection at the chamber. It also lets us diagnose many problems without a site visit.

Nitrogen cryosaunas: two extra requirements

Oxygen monitoring and ventilation

Liquid nitrogen displaces oxygen. A nitrogen cryosauna room needs a working oxygen monitor and good mechanical ventilation.

  • Normal air contains about 20.9% oxygen. OSHA treats air below 19.5% oxygen as oxygen-deficient and dangerous.
  • If the oxygen alarm sounds: stop the session, get everyone out and ventilate the room. Don’t go back in until the oxygen monitor reads at least 19.5%, ideally back to about 20.9%.
  • If someone feels dizzy, light-headed, short of breath or confused, get them into fresh air and call 911.
  • Never reset, bypass or silence the alarm to continue sessions.

Oxygen sensor alarm: what to do →

Nitrogen storage and delivery

  • A safe place for the tank, as specified by your gas supplier and the manufacturer.
  • Delivery access for regular tank exchanges.
  • Room size and ventilation to the manufacturer’s specification. One manufacturer’s manual, for example, asks for a room of at least 10–15 m² (about 110–160 sq ft) with a ceiling of at least 250 cm (about 8.2 ft) and efficient mechanical ventilation.

Room checklist

Before delivery:

  • ☐ Delivery path measured: doors, corridors, turns, elevator
  • ☐ Floor load confirmed for the chamber’s weight
  • ☐ Installation space and ceiling height match the manufacturer’s spec
  • ☐ Service access behind and around the chamber
  • ☐ Dedicated circuit installed to spec (single-phase or 3-phase)
  • ☐ Heat removal planned: room AC sized, or condenser location and line-set route approved
  • ☐ Permits checked (electrical, mechanical)
  • ☐ Dehumidifier in place; room below 40% humidity
  • ☐ Ventilation meets the manufacturer’s air-change requirement
  • ☐ Ethernet or Wi-Fi at the chamber

Nitrogen cryosaunas, also:

  • ☐ Oxygen monitor installed and tested
  • ☐ Mechanical ventilation to spec
  • ☐ Tank location and delivery access agreed with the gas supplier

Frequently asked questions

Does a cryotherapy chamber need 3-phase power?
Many first-generation electric chambers do. Newer self-contained chambers run on single-phase 208–240 V. Check the manufacturer’s specification before you sign a lease or order a chamber.

What humidity should a cryotherapy room have?
Below 40% relative humidity. Higher humidity leads to ice on fans, lifts and sensors, which is the most common cause of avoidable repairs we see.

Does a cryotherapy chamber need an outdoor condenser?
Split-system chambers do. Self-contained chambers release their heat into the room instead, so the room needs enough air conditioning.

How much space does a cryotherapy chamber need?
It depends on the model. Plan for the manufacturer’s installation space, which is larger than the chamber itself, plus room for service access. A single-person chamber typically needs a room of around 8 × 12 ft.

What does a nitrogen cryosauna room need?
A working oxygen monitor, mechanical ventilation, and a safe place and delivery route for the nitrogen tank.

Planning an installation?

We install and relocate cryotherapy chambers nationwide. Send us your room dimensions, power supply and photos of the space, and we’ll tell you what needs to change before delivery. Installation service → · Call (213) 559-0665